Radiated Field TEST SETUP From: 61000-4-3 . IEC:2002+A1:2002
 
7 Test set-up

All testing of equipment shall be performed in a configuration as close as possible to the installed case. Wiring shall be consistent with the manufacturer's recommended procedures, and the equipment shall be in its housing with all covers and access panels in place, unless otherwise stated.

If the equipment is designed to be mounted in a panel, rack or cabinet, it shall be tested in this configuration.

A metallic ground plane is not required. When a means is required to support the test sample, it shall be constructed of a non-metallic, non-conducting material. However, grounding of housing or case of the equipment shall be consistent with the manufacturer's installation recommendations.

When an EUT consists of floor-standing and table-top components, the correct relative positions shall be maintained.

Typical EUT set-ups are shown in figures 5 and 6.
 
rad
 
rad
 
7.1 Arrangement of table-top equipment

The equipment to be tested is placed in the test facility on a non-conducting table 0,8 m high.

NOTE The use of non-conducting supports prevents accidental earthing of the EUT and distortion of the field. To ensure the latter, the support should be bulk non-conducting, rather than an insulating coating on a metallic structure.

The equipment is then connected to power and signal wires according to relevant installation instructions.
 
7.2 Arrangement of floor-standing equipment

Floor-standing equipment shall be mounted on a non-conducting support 0,1 m above the supporting plane. The use of non-conducting supports prevents accidental earthing of the EUT and distortion of the field. To ensure the latter, the support shall be bulk non-conducting, rather than an insulating coating on a metallic structure. Floor-standing equipment which is capable of being stood on a non-conducting 0,8 m high platform, i.e. equipment which is not too large or heavy, or where its elevation would not create a safety hazard, may be so arranged, if specifically required by the product committees. This variation in the standard method of test shall be recorded in the test report.

The equipment is then connected to power and signal wires according to relevant installation instructions.
 
7.3 Arrangement of wiring

If the wiring to and from the EUT is not specified, unshielded parallel conductors shall be used.

Wiring is left exposed to the electromagnetic field for a distance of 1 m from the EUT.

Wiring between enclosures of the EUT shall be treated as follows :

the manufacturer's specified wiring types and connectors shall be used;
   
if the manufacturer's specification requires a wiring length of less than or equal to 3 m, then the specified length shall be used. The wiring shall be bundled low-inductively to 1 m length;
   
if the specified length is greater than 3 m, or is not specified, then the illuminated length shall be 1 m. The remainder is decoupled, for instance via lossy r.f. ferrite tubes.


The EMI filtering used shall not impair the operation of the EUT. The method used shall be recorded in the test report.

In one EUT position, the wires shall be arranged parallel to the uniform area of the field to minimize immunity.

All results shall be accompanied by a complete description of the wiring and equipment position and orientation so that results can be repeated.

The bundled length of exposed wiring is run in a configuration which essentially simulates normal wiring; that is, the wiring is run to the side of the EUT, then either up or down as specified in the installation instructions. The horizontal/vertical arrangement helps to ensure worst-case conditions.

 
7.4 Arrangement of human body-mounted equipment

IHuman body-mounted equipment may be tested in the same manner as table top items. However, this may involve over-testing or under-testing because the characteristics of the human body are not taken into account. For this reason, product committees are encouraged to specify the use of a human body simulator with appropriate dielectric characteristics.